Use these steps to ensure rotation is working in your SmartLabel device before running an experiment.
- Click on the rotation speed. This will bring you into the rotation settings window.
- Click the button that displays the current speed of 0.01.
- On the right, click and drag the scroll bar to a speed between 2 and 3 RPM.
- Press “save” then “escape”.
- Start rotation.
- Look to see if the cup is spinning smoothly in the chamber. Make sure the cup can do one full rotation.
- If the cup is struggling to rotate, refer to the Troubleshooting guide‘s section on rotation.
- If the cup rotates successfully, open the rotation window again.
- Click on the button that shows the current speed.
- Click and drag the bar all the way to the bottom so it shows 0 RPM.
- Press the up arrow once so the speed is 0.01 RPM.
- Save.
- Escape.
Make sure rotation is set to 0.01 RPM before beginning any experiments.
- Turn rotation on again.