Use these steps to ensure rotation is working in your SmartLabel device before running an experiment.

  • Click on the rotation speed. This will bring you into the rotation settings window.
  • Click the button that displays the current speed of 0.01.
  • On the right, click and drag the scroll bar to a speed between 2 and 3 RPM.
  • Press “save” then “escape”.
  • Start rotation.
  • Look to see if the cup is spinning smoothly in the chamber. Make sure the cup can do one full rotation.
  • If the cup rotates successfully, open the rotation window again.
  • Click on the button that shows the current speed.
  • Click and drag the bar all the way to the bottom so it shows 0 RPM.
  • Press the up arrow once so the speed is 0.01 RPM.
  • Save.
  • Escape.

Make sure rotation is set to 0.01 RPM before beginning any experiments.

  • Turn rotation on again.